Low radiation-dose CT Colonography

This study further validates that model-based iterative reconstruction can decrease CT exam dose by 50-80% without compromising diagnostic power. There now is a substantial accumulation of published reports of this type in multiple body areas and organs. The same is becoming true for blended types of adaptive statistical plus model based (minus the optical components) iterative reconstruction (such as ASIR-V).

CT colonography

Using model-based iterative reconstruction, CT colonography can be a very low radiation dose method of screening. This article applauds the United States Preventive Services Task Force (USPSTF) approval, cited as a “big win for patients.”

CT Colonography: Reducing the Radiation Dose

This interesting paper talks about the use of iterative reconstruction to help lower the radiation dose of screening CT colonography.

Of course, as with all screening exams, the first order of priorities is to do no harm – hence the motivation to keep the radiation dose especially low.

The challenge is to lower dose without compromising diagnostic power.

For about the past two years, here at UW Medicine (Seattle) we have been using Model Based Iterative Reconstruction (VEO, GE Healthcare) for all our CT colonography exams. As recommended in this article, we also keep the kVp low – 80 or 100, which also helps to reduce the dose.

The result is a very low dose exam, but with excellent image quality and low image noise. This helps to make great coronal/sagittal reconstructions plus very nice 3D fly-through on the post-processing workstation.